Turbidity Meter Series WTW Turb PLUS 2000

Descripción general Especificaciones
Descripción general

Descripción general

Standard-compliant turbidity measurement with ultrasonic cleaning, according to US EPA 180.1 and DIN EN ISO 7027.

Nephelometric turbidity meter with white (US EPA 180.1) or infrared light (EN ISO 7027), with ultrasonic cleaning.

Características del producto
  • Safe measurements by integrated bubble trap and ultrasonic cleaning
  • Easy calibration with standards that are traceable to Formazine
  • Standard-compliant measurement according to US EPA 180.1 and DIN EN ISO 7027, respectively
  • High accuracy, especially below 40 NTU

The nephelometric turbidity meters of series Turb PLUS 2000 can be used for nearly all applications thanks to its large measuring range. By using white or infrared light and the respective standards, the meters meet the requirements of the US EPA 180.1 as well as those of the DIN EN ISO 7027. Instrument with white as well as infrared light are equipped by default with an integrated bubble trap and ultrasonic cleaning, which saves time and effort. If needed, a set with additional external bubble trap is available.

Calibration solutions need to be ordered separately.

Turb PLUS 2020:
Online turbidity meter, with white light and integrated bubble trap; nephelometric measurement specified
according to US EPA 180.1, with ultrasonic cleaning 110–240 VAC.

Turb PLUS 2120:
Online turbidity meter, with infrared light and integrated bubble trap; nephelometric measurement
specified according to DIN EN ISO 7027, with ultrasonic cleaning 110–240 VAC.

Turb PLUS 2120 Set:
Turb PLUS 2120 including external bubble trap.

Especificaciones

Especificaciones

Specifications Turb PLUS 2020 Turb PLUS 2120
Measurement principle Scattered light measurement, white light, nephelometric measurement Scattered light measurement, white light, nephelometric measurement
Measurement Range Factory setting: 0 ... 100 NTU; optional measuring ranges 0 ... 10 and 0 ... 1000 NTU (loaded on USB) Factory setting: 0 ... 100 NTU; optional measuring ranges 0 ... 10 and 0 ... 1000 NTU (loaded on USB)
Resolution Down to 0.0001 NTU Down to 0.0001 NTU
Accuracy ±2 % of reading or ±0.02 NTU below 40 NTU whichever is greater (0–100 and 0–1000 ranges) ±5 % of reading above 40 NTU (0–100 and 0–1000 NTU ranges) ±2 % of reading or ±0.02 NTU (0–10 NTU range) ±2 % of reading or ±0.02 NTU below 40 NTU whichever is greater (0–100 and 0–1000 ranges) ±5 % of reading above 40 NTU (0–100 and 0–1000 NTU ranges) ±2 % of reading or ±0.02 NTU (0–10 NTU range)
Limit of Detection 0.02 mg/l 0.02 mg/l
Response time Adjustable (1 ... 60 readings) Adjustable (1 ... 60 readings)
Sample Temperature 1 °C ... 50 °C (34 °F ... 122 °F) 1 °C ... 50 °C (34 °F ... 122 °F)
Sample Flow Rate 100 ml/min. ... 1.5 liter/min. (0.026 0.40 Gal/min) 100 ml/min. ... 1.5 liter/min. (0.026 0.40 Gal/min)
Maximum Water Pressure Integral pressure regulator rated 7 bar (101.5 PSI) Integral pressure regulator rated 7 bar (101.5 PSI)
Cleaning Ultrasonic cleaning Ultrasonic cleaning
Calibration Factory calibrated to 0 100 NTU, recalibration required every three months and after change of measuring range Factory calibrated to 0 100 NTU, recalibration required every three months and after change of measuring range
Display Multi-Line Liquid Crystal Backlit Display Multi-Line Liquid Crystal Backlit Display
Mounting Wall mounting with 6 screws Wall mounting with 6 screws
Regulatory Compliance Compliant to U.S. EPA 180.1 Compliant to DIN EN ISO 7027 Compliant to U.S. EPA 180.1 Compliant to DIN EN ISO 7027
Power 100–240 V AC, 47–63 Hz, 80 VA 100–240 V AC, 47–63 Hz, 80 VA
Relays Two Programmable, 120–240 V AC 2A Form C Relay Two Programmable, 120–240 V AC 2A Form C Relay
Analog Output One powered 4–20 mA, 1000 W drive One powered 4–20 mA, 1000 W drive
Digital Output Bi-directional RS-485, Modbus RTU Bi-directional RS-485, Modbus RTU
Environmental conditions Storage Temperature Range: 1 °C ... 50 °C (34 °F 122 °F), Operating Temperature Range: 1 °C ... 50 °C (34 °F 122 °F), Relative Humidity: Up to 95 % (non-condensing) Not recommended for outdoor use. Altitude up to 2000 meters. Storage Temperature Range: 1 °C ... 50 °C (34 °F 122 °F), Operating Temperature Range: 1 °C ... 50 °C (34 °F 122 °F), Relative Humidity: Up to 95 % (non-condensing) Not recommended for outdoor use. Altitude up to 2000 meters.
Certification CE, cETLus, CE Approved, LC listed to UL 61010-1 3rd Edition: 2012 LC: Certified to CSA 22.2 No.61010-1-12: 2012 CE, cETLus, CE Approved, LC listed to UL 61010-1 3rd Edition: 2012 LC: Certified to CSA 22.2 No.61010-1-12: 2012
Mechanics Wetted Materials: Nylon, Wetted Materials: Nylon, Borosilicate Glass, Silicone, Polypropylene, Stainless Steel, Viton, Acety Housing: ABS (Double Insulated, Pollution Degree 2, Overvoltage Category II) Wetted Materials: Nylon, Wetted Materials: Nylon, Borosilicate Glass, Silicone, Polypropylene, Stainless Steel, Viton, Acety Housing: ABS (Double Insulated, Pollution Degree 2, Overvoltage Category II)
Weight 2.8 kg (6.2 lbs.) 2.8 kg (6.2 lbs.)
Warranty 2 years on defects in quality according to § 10 terms of conditions 2 years on defects in quality according to § 10 terms of conditions