Order Number
|
201650
|
201652
|
201660
|
201653
|
Measuring method
|
Optical
|
Optical
|
Optical
|
Optical
|
Measuring range (25 °C)
|
-
|
-
|
-
|
-
|
O2 concentration
|
0 ... 20.00 mg/l (0 ... 20.00 ppm)
|
0 ... 20.00 mg/l (0 ... 20.00 ppm)
|
0 ... 20.00 mg/l (0 ... 20.00 ppm)
|
0 ... 20.00 mg/l (0 ... 20.00 ppm)
|
O2 saturation
|
0 ... 200.0 %
|
0 ... 200.0 %
|
0 ... 200.0 %
|
0 ... 200.0 %
|
Resolution
|
-
|
-
|
-
|
-
|
O2 concentration
|
0.01 mg/l (0.01 ppm)
|
0.01 mg/l (0.01 ppm)
|
0.01 mg/l (0.01 ppm)
|
0.01 mg/l (0.01 ppm)
|
O2 saturation
|
0.1 %
|
0.1 %
|
0.1 %
|
0.1 %
|
Accuracy
|
< 1 mg/l (ppm): ± 0.05 mg/l (ppm) > 1mg/l (ppm): ± 0.1 mg/l (ppm)
|
< 1 mg/l (ppm): ± 0.05 mg/l (ppm) > 1mg/l (ppm): ± 0.1 mg/l (ppm)
|
< 1 mg/l (ppm): ± 0.05 mg/l (ppm) > 1mg/l (ppm): ± 0.1 mg/l (ppm)
|
< 1 mg/l (ppm): ± 0.05 mg/l (ppm) > 1mg/l (ppm): ± 0.1 mg/l (ppm)
|
Response time at 25 °C
|
t90: < 150 s t95: < 200 s
|
t90: < 150 s t95: < 200 s
|
t90: < 60 s t95: < 80 s
|
t90: < 60 s t95: < 80 s
|
Minimum flow rate
|
No flow required
|
No flow required
|
No flow required
|
No flow required
|
SensCheck
|
Monitoring of membrane function
|
Monitoring of membrane function
|
Monitoring of membrane function
|
Monitoring of membrane function
|
Temp. measurement
|
Integrated NTC, 23 °F ... 140 °F (-5 °C ... +60 °C) ± 0.5 °C
|
Integrated NTC, 23 °F ... 140 °F (-5 °C ... +60 °C) ± 0.5 °C
|
Integrated NTC, 23 °F ... 140 °F (-5 °C ... +60 °C) ± 0.5 °C
|
Integrated NTC, 23 °F ... 140 °F (-5 °C ... +60 °C) ± 0.5 °C
|
Temp. compensation
|
23 °F ... 122 °F (-5 °C ... +50 °C)
|
23 °F ... 122 °F (-5 °C ... +50 °C)
|
23 °F ... 122 °F (-5 °C ... +50 °C)
|
23 °F ... 122 °F (-5 °C ... +50 °C)
|
Pressure Resistance
|
10 bar (incl. sensor connection cable)
|
10 bar (incl. sensor connection cable)
|
10 bar (incl. sensor connection cable)
|
10 bar (incl. sensor connection cable)
|
Ambient Conditions
|
23 °F ... 122 °F (-5 °C ... +50 °C) -13 °F ... 122 °F (-25 °C ... +50 °C)
|
23 °F ... 122 °F (-5 °C ... +50 °C) -13 °F ... 122 °F (-25 °C ... +50 °C)
|
23 °F ... 104 °F (-5 °C ... +40 °C) -13 °F ... 104 °F (-25 °C ... +40 °C)
|
23 °F ... 104 °F (-5 °C ... +40 °C) -13 °F ... 104 °F (-25 °C ... +40 °C)
|
Electrical connections
|
2-wire shield cable with quick fastener to sensor
|
2-wire shield cable with quick fastener to sensor
|
2-wire shield cable with quick fastener to sensor
|
2-wire shield cable with quick fastener to sensor
|
Electromagnetic Compatibility
|
EN 61326, Class B, FCC Class A; Intended for indispensable operation
|
EN 61326, Class B, FCC Class A; Intended for indispensable operation
|
EN 61326, Class B, FCC Class A; Intended for indispensable operation
|
EN 61326, Class B, FCC Class A; Intended for indispensable operation
|
Certifications
|
CE, cETL, ETL
|
CE, cETL, ETL
|
CE, cETL, ETL
|
CE, cETL, ETL
|
Mechanical
|
Sensor cap, fixation: POM, PVC, silicone, PMMA sensor body: VA stainless steel 1.4571 protection type IP 68
|
Sensor cap, fixation: POM, PVC, silicone, PMMA sensor body: VA stainless steel 1.4571 protection type IP 68
|
Sensor cap, fixation: POM, PVC, silicone, PMMA sensor body: VA stainless steel 1.4571 protection type IP 68
|
Sensor cap, fixation: POM, PVC, silicone, PMMA sensor body: VA stainless steel 1.4571 protection type IP 68
|
Weight (without cable)
|
1.98 lb (900 g)
|
3.31 lb (1.5 kg)
|
1.98 lb (900 g)
|
3.31 lb (1.5 kg)
|
Warranty
|
2 years for defects in quality
|
2 years for defects in quality
|
2 years for defects in quality
|
2 years for defects in quality
|